Diagnostics
ISI Dos version
Enter
in scan and then activate the EntryLevel at level 2:


Now under Diagnostics you will see the menu of
options:
In
the repeatability options set Number of
repeated complete scans to display equal to 10 and also set Output to printer to YES.

Set
Output to printer to YES under the Instrument Response options and the Wavelength accuracy options. In the Wavelength accuracy options Number of
repeated complete scans to display equal to 5.
The
printed output should look like the one on the next page:

Instrument Response on 08-28-2000
Filter Minimum Data
Maximum Data
Number Value Point
Value Point
1 2844 2
55313 143
Dark Period Values
1 862
2 864
_
Photometric Repeatability 16 revs 10 scans 08-28-2000
Scan RMS BIAS
RMS(C)
1 16 -12
10
2 20 -17
12
3 11 -3
10
4 16 10
12
5 11 -5 10
6 19 -14
12
7 34 -32
12
8 11 -3
10
9 29 -27
9
10 21 -18
11
Minimum
11 -32 9
Maximum
34 10 12

Average
20 -12 11
_
Scan
Wavelength Reference Peaks
08-28-2000
1 1143.83
1680.96 2166.64 2305.91
2 1143.82
1680.96 2166.64 2305.91
3 1143.82
1680.96 2166.64 2305.91
4 1143.82
1680.96 2166.64 2305.91
5 1143.82 1680.96 2166.64
2305.91
Peak Average Diff S.D.
1143.63 1143.82 0.1922
0.0046
1681.27 1680.96 -0.3138
0.0011
2166.40 2166.64 0.2418
0.0015
2305.93 2305.91 -0.0208
0.0008
Current Suggested
K
3311.02 3311.07
Phi
0.31823 0.31821
Error 0.2204 0.2194