Date: 3-29-01
NEW Check Cell installation:
NOTE: Replacement of the old with the new check cell should be performed only for the labs that have been standardized. For those few labs that have not been standardized in the past year yet, the procedure should be performed AFTER the standardization.
1.
Install
the ‘revised’ check cell equation:
a.
DOS
: under Equation management

Modify the equation name under Standard check cell. In the example change ISISTDF2.eqa with the ISISTDF4.eqa equation in the diskette.

B. Windows. Click on
‘analysis’, then on
‘files’, and
‘advanced’, click on
and type ‘ISI check cell equation. Under
‘Model Setup’ (see figure) select the new equation “ISIstdf4.eqa” and also
check on the ‘Check cell’.
![]()
![]()

2.
Make
sure you are using the new standardization during scanning:
a.
DOS.
In this example the std file is ‘auto0799.std’. You should use the one received
from Paolo.

b.
Windows. Click on
‘analysis’, then on
‘files’, and
‘advanced’, under ‘data collection’ you will
find ‘Standardization filename’; In this example the std file is
‘auto0799.std’. You should use the one received from Paolo.
![]()

3.
Scan
the cell 3 times. The program will automatically create a file called
‘Isistdf4.chk’;
4.
Send
the ‘Isistdf4.chk’ file to Paolo (‘pberzaghi@facstaff.wisc.edu’) along with the
equation file (‘isistdf4.eqa’) from the diskette;
5.
Paolo
will evaluate the spectra and if necessary he will make changes to the
equation;
6.
For
any problem call (ph 608-264-5232) or email Paolo.