Instrument Monitoring (Example 1)

 

 

Lab name:____________                          Date of submission: 4-17-00

                                                                  Date of the last test: 4-17-00

 

 

 

Summary

Response Test               Pass

Wavelength Accuracy     Pass

Repeatability                  Pass

 

Your instrument has passed all the tests please remember to rerun diagnostics by 4-24-00 (Date of the last day+7).

 

See details of the test


Diagnostic Details

 

 

Actual

Limits

Pass/Fail

Response Test

 

 

 

         Maximum Value

32658

Greater then 30000

         Dark Value

1067

Lower then 1600

 

 

 

 

Wavelength Accuracy

 

 

 

         Error

0.237

Lower than 0.300

Repeatability

 

 

 

         RMS©

 

 

 

                   Visible

17

Lower than 20

                   NIR

17

Lower than 20

         Bias

 

 

 

                   Visible

-11

Lower than 10

                   NIR

-14

Lower than 10

 


Wavelength Accuracy

 

Peak

Average

Difference

S.D.

 

 

Within ± 1.0

Lower than 0.020

1143.63

1143.66

0.0289

0.0010

1681.27

1680.98

-0.2877

0.0015

2166.40

2166.49

0.0925

0.0023

2305.93

2305.78

-0.1507

0.0030

 

 

 

 

 

Actual Values

Suggested Values

K

3305.53

3305.64

PHI

0.32296

0.32296

 

Lower than 0.3

 

Error

0.169

0.152