Instrument Monitoring (Example 1)
Lab
name:____________ Date
of submission: 4-17-00
Response Test Pass
Wavelength Accuracy Pass
Repeatability Pass
Your instrument has passed all the tests please
remember to rerun diagnostics by 4-24-00 (Date of the last day+7).
|
|
Actual |
Limits |
Pass/Fail |
|
Response Test |
|
|
|
|
Maximum
Value |
32658 |
Greater then 30000 |
|
|
Dark
Value |
1067 |
Lower then 1600 |
|
|
|
|
|
|
|
|
|
|
|
|
Error |
0.237 |
Lower than 0.300 |
|
|
Repeatability |
|
|
|
|
RMS© |
|
|
|
|
Visible |
17 |
Lower than 20 |
|
|
NIR |
17 |
Lower than 20 |
|
|
Bias |
|
|
|
|
Visible |
-11 |
Lower than 10 |
|
|
NIR |
-14 |
Lower than 10 |
|
|
Peak |
Average |
Difference |
S.D. |
|
|
|
Within ± 1.0 |
Lower than 0.020 |
|
1143.63 |
1143.66 |
0.0289 |
0.0010 |
|
1681.27 |
1680.98 |
-0.2877 |
0.0015 |
|
2166.40 |
2166.49 |
0.0925 |
0.0023 |
|
2305.93 |
2305.78 |
-0.1507 |
0.0030 |
|
|
Actual Values |
Suggested Values |
|
K |
3305.53 |
3305.64 |
|
PHI |
0.32296 |
0.32296 |
|
|
Lower than 0.3 |
|
|
Error |
0.169 |
0.152 |